HALO TECHNOLOGY9,10,11,12,13
The proprietary technology incorporated in the HALO360 and HALO90 Systems is designed to maximize clinical outcomes and completely remove the Barrett's epithelium without significant injury to the underlying tissue.
 |
| Bipolar electrode array geometry precisely controls the depth of ablation to < 1 mm. |
- Delivery of ablative energy in less than 1 second allows long or short segments of Barrett's to be treated quickly
- Consistent application of bipolar energy uniformly removes the esophageal epithelium, reducing potential for buried
glands and improving patient tolerability
- Controlled treatment depth of less than 1,000 µm reduces risk of stricture formation, even after multiple energy applications
PRECISE DEPTH CONTROL
Barrett's epithelium is approximately 500 µm in thickness. The HALO Energy Generator and the HALO Ablation Catheter electrode
array are designed to work in concert to achieve a uniform, superficial depth of ablation of ~1,000 µm.
Human Esophagus Specimen (H&E Stain)
PREDICTABLE OUTCOME
In both porcine and human esophagectomy studies, the HALO Technology has demonstrated
removal of the esophageal epithelium without significant injury to the submucosa
when using the recommended treatment settings.
| |
|
|
|
| Non-ablated esophagus Epithelium intact |
|
48 hours Post-Ablation Absence of epithelium with preservation of lamina propria and muscularis mucosae. |